Edwards Ion gauge IG40 BA DN40CF
    search
    • Edwards Ion gauge IG40 BA DN40CF

    Edwards Ion gauge IG40 BA DN40CF

    HVC-D03000300

    Ion gauge IG40 BA DN40CF

    The Edwards range of passive gauges comprises of Pirani, Penning and both Bayard-Alpert and Extractor Ion gauge heads which together can measure from atmosphere down to 10-12 mbar. These are used in conjunction with a Passive gauge controller (PGC) to display pressure and offer an intuitive interface for the user alongside the ability to control remotely. Edwards will offer 2 controller variants; the PGC201 which covers the Pressure range of 10-9 to 1000 mbar, in conjunction with Pirani and Penning gauges, and the PGC202 which covers the wider range of and 10-12 to 1000 mbar when used with a Pirani and Ion gauge.


    Price on request

    Offerte aanvragen

    Technical Data

    UnitsIG40 BAIG40 EX
    Measurement rangembar (Torr)2 x 10-11 to 10-2
    (1.5 x 10-11 to 10-2)
    2x10-12 to 10-4
    (1.5 x 10-12 to 10-4)
    X-ray limitmbar (Torr)≤10-11 (≤10-11)≤10-12 (≤10-12)
    Ambient temperature during operation°C20 to +8020 to +80
    Maximum flange temperature with
    bakeable gauge cable
    °C250250
    Maximum bakeout temperature
    (with no cable connected)
    °C400400
    Material
    Cathode
    Anode
    Collector
    Reflector
    Iridium with yttrium oxide coating
    Pt/Ir 90/10 and Mo/pt wrapped wire
    Tungsten
    Iridium with yttrium oxide coating
    Mo and CoNiCr
    Tungsten
    NiFe
    Vacuum connectionDN40CFDN40CF
    Operating characteristics
    Ion detector potential
    Cathode potential
    Anode potential
    Reflector potential
    V0
    80
    220
    -
    0
    100
    220
    205
    Emission currentmA0.1 to 10.01.6
    Heating current for the hot cathodeA1.51.5
    Heating voltage for the hot cathodeV3.03.7
    Sensitivity for nitrogenmbar-117.06.25
    Bake out operation, Electron bombardmentV/mA480/90480/45
    Compatible controllersPGC202PGC202

    Dimensions

    Image.ashx?id=352925

    Bayard-Alpert sensing system

    • Measurement range from 10-2 to 2 x 10-11 mbar (1.5 x 10-11 Torr)
    • Protection shield welded in place

    Extractor sensing system
    • Measurement range from 10-4 to 2 x 10-12 mbar (1.5 x 10-12 Torr)
    • Significant reduction of X-ray and ion desorption effects
    Type
    Ion Gauge
    Measuring range
    2 · 10-11 – 10-2 hPa
    Gauge model
    IG40 BA
    Flange