D14641000 AIM-S-NW25
    search
    • D14641000 AIM-S-NW25

    AIM-S-NW25

    HVC-D14641000

    The AIM Gauge is an inverted magnetron gauge head and gauge controller in a single compact unit. The gauge operates as a cold cathode ionisation gauge, in which pressure is measured indirectly as a function of the current that flows in a Townsend discharge maintained in the body tube.


    The measurement range of the AIM Gauge is 1 x 10-8 to 1 x 10-2 mbar.

    The AIM Gauge requires a 13.5 to 36 V d.c. power supply: it has a 2 to 10 V d.c. analogue output which is related to pressure.

    The AIM Gauge is compatible with all of Edwards Active Gauge Controllers and with the appropriate versions of the Active Gauge Displays. Alternatively you can use an independent power supply for the AIM Gauge and you can read the AIM Gauge output signal with a voltmeter or an analogue-to-digital converter.

    € 1,375.00
    (tax excl.)
    Not on stock

    Request quote
    Pressure Range
    AIM-S10-2 to 10-8 mbar
    AIM-X10-2 to 10-9 mbar
    Accuracy*Typically ±30%
    Maximum overpressure10 bar absolute (145 psi)
    Power supply+13.5 to +36 V DC (max 1 V ripple)
    Power consumption2 W Maximum
    Output signal2 to 10 V DC
    Set pointOpen collector transistor
    Maximum voltage40V DC
    Current100 mA max
    Temperature range
    Operating+5 to +60°C
    Storage+0 to +70°C
    Materials exposed to vacuum
    NW25 versionsStainless steel 304 & 316 & 347, fluoroelastomer, glass
    DN40CF versionsStainless steel 304 & 316 & 347, copper, ceramic
    Internal volume26 cm3
    Weight0.81 kg
    External interface connector8-way FCC68/RJ45 socket
    Vacuum fittingNW25
    Standards
    Electronic designEN 61010-1
    Electromagnetic compatibilityEN 61326 (Class B Emissions)
    Flame retardant casingUL94
    Enclosure ratingIP40

    *Accuracy is reduced at the limits of the measuring range

    Performance

    Image.ashx?id=107477

    Dimensions

    Image.ashx?id=107476
    Type
    Active Inverted Magnetron gauge
    Measuring range
    10-2 to 10-8 mbar
    Gauge model
    AIM-S
    Precision: % of measurement [%]
    Typically ±30%
    Flange
    DN 25 ISO-KF